热门关键词:苏玛罐,原子吸收分光光度计,环境监测产品,液相色谱,色谱柱。
更新时间:2019-11-20
IonPac Fast Anion IIIA阴离子交换柱062964这种坚固的氢氧化物选择性阴离子交换柱提供快速,可靠的结果,采用等度洗脱液通过氢氧化钾与抑制电导检测相结合的淋洗液发生器交付。
免费咨询:010-61755520
分析磷酸和柠檬酸与Thermo Scientific? Dionex? IonPac? Fast Anion IIIA柱在不到7分钟,在可乐软饮料样品的酸。这种坚固的氢氧化物选择性阴离子交换柱提供快速,可靠的结果,采用等度洗脱液通过氢氧化钾与抑制电导检测相结合的淋洗液发生器交付。
IonPac Fast Anion IIIA阴离子交换柱062964
订货信息:
分析柱 | |
IonPac Fast Anion IIIA Analytical Column (3 x 250 mm) | 062964 |
保护柱 | |
IonPac Fast Anion IIIA Guard Column (3 x 50 mm) | 062966 |
This hydroxide-selective anion-exchange column is specifically designed for the determination of phosphoric and citric acids in cola soft drinks. Its capacity and selectivity allows rapid analysis of these acids in less than 7 minutes.
Use with the Eluent Generator for simplified Reagent-Free IC operation.
The 3 mm i.d. configuration provides economical operation.
Combine with the ASRS? 300 suppressor for low background and enhanced sensitivity.
Compatible with organic solvents.
The Fast Anion IIIA is optimized for a 30 °C operating temperature to ensure reproducible retention times in all environmental conditions. It is compatible with organic solvents to enhance analyte solubility, modify column selectivity, or allow effective column cleanup. The Fast Anion IIIA column is recommended for use with the Eluent Generator, requiring only a deionized water source to produce potassium hydroxide eluent.
The cola samples may require sample pretreatment with sonication to remove carbonation. Sample dilution of cola syrups is recommended prior to analysis to ensure optimum column life. These acids can be analyzed in less than 7 minutes using an isocratic potassium hydroxide eluent delivered by an Eluent Generator in combination with suppressed conductivity detection.
IonPac Fast Anion IIIA Anion-Exchange Column Specifications | |
---|---|
Dimensions | IonPac Fast Anion IIIA Analytical Column: 3 × 250 mm IonPac Fast Anion IIIA Guard Column: 3 × 50 mm |
Mobile phase compatibility | pH 0–14; 0–100% HPLC solvents |
Ion-exchange group | Functional Group: Alkanol quaternary ammonium ion |
Maximum operating pressure | 3000 psi |
Functional group characteristics | Hydrophobicity: Ultralow hydrophobic |
Capacity | 55 μeq (3 × 250 mm analytical column) 1.0 μeq (3 × 50 mm guard column) |
Column construction | PEEK with 10-32 threaded ferrule style end fittings. All components are nonmetallic. |
Substrate characteristics | Bead Diameter and Pore Size: 7.5 μm (3 × 250 mm); supermacroporous resin (2000 ?) 11 μm (3 × 50 mm); microporous resin (<10 ?) Cross-Linking (%DVB): 55% |